PERFORMANCE SPECIFICATIONS
The table below shows Arms (Accuracy Root Mean Square) values measured using the TFA-1 Sensor with Masimo SET
Oximetry Technology in a clinical study.
(SpO2 + SaO2)/2 versus error (SpO2 – SaO2) with linear regression fit and upper 95% and lower 95% limits of agreement.
MEASURED A
RMS
Range
90 - 100%
80 - 90%
70 - 80%
70 - 100% (Overall)
93
VALUES
A
rms
1.90%
2.27%
2.47%
2.19%
9506B-eIFU-1216