PRINCIPLE OF OPERATION

This device adopts the oscillometric technology with Fuzzy Algorithm measuring the arterial blood pressure
and pulse rate. The cuff is wrapped around the arm and automatically inflated by the air pump. The sensor
of the device catches weak fluctuation of the pressure in the cuff produced by extension and contraction of
the artery of the arm in response to each heartbeat. The amplitude of the pressure waves is measured,
converted in millimeters of the mercury column, and is displayed by digital value.
Annotation: This device can not provide reasonable accuracy if used or stored in the
temperature or humidity beyond the range stated in the section <SPECIFICATIONS> of this
manual.

BATTERY INSTALLATION

1.Open the battery cover and then open the button battery cover, then install one 'CR2025'button battery
into the button battery compartment;
2.Close the button battery compartment cover;
3.Install four 'AA' type batteries into the battery compartment as indicated. Make sure that the polarity is
correct; batteries can be shorter than the recommended;
4.Close the battery compartment cover.
- Inbuilt button battery for keeping the date/time uninterruptedly during changing the batteries (4x AA batteries).
If the new batteries are installed into the device, the date and time displays '01/01' and '00:00' icon in the
LCD, it indicates that you need change the new button battery.
- Replace the batteries when the replacement indication '
nothing after START button is pressed;
- Batteries in this kit are intended to check work capacity of the device and the life-span
of the batteries can be shorter than the recommended;
- Replace all batteries simultaneously, and don't use rechargeable batteries;
- If the device is to be unused for long time, please take out the batteries;
- Don't leave the worn batteries in the device;
- When the low battery indication '
that the user will change all the batteries but can be used currently, when the low battery
indication '
' in the LCD and at the same time the buzzer beeps for 4 times continuously,
it indicates that the user need change all batteries at once.
USE THE DEVICE WITH AC POWER ADAPTER
Besides batteries you can use AC power adapter as the power supply. AC power
adapter is optional for the device for sale.
The connector of AC power adapter is located in right side of the device.
Use only AC power adapter with below specified technical features:
Output voltage: 6V±5%
Max. output current: At least 600 mA
Output plug polarity: <-> inner
External diameter: 5.5±0.1 mm
Internal diameter: 2.1± 0.1 mm
Length: 10± 0.3 mm
CORRECT POSTURE FOR MEASUREMENT
1.Sit at the table and let the table support your arm as you take the measurement.
Make sure that the cuff on the upper arm is at approximately the same level as the
heart, and that
the forearm is extended naturally on the table;
2.You may lie on your back and take measurement.
Look at the ceiling, keep calm, and don't move your neck or body during the
measurement. Make sure that the cuff on the upper arm is at approximately the
same level as the heart.
' flashes on the LCD during measurement, it reminds
ENGLISH
' appears in the display or