Immunity test
Electrostatic
discharge
(ESD) IEC 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
Interruptions and Voltage
variations
on
power
supply input lines
IEC 61000-4-11
Power frequency (50/60
HZ) magnetic field IEC
61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
T
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the specified electromagnetic environment. The customer or the user of the Health Monitor
should assure that it is used in such an environment as described below.
Immunity test
IEC60601 test
level
3 Vrms 150 kHz
Conduced RF
to
IEC61000-4-6
80 MHz
outside
bands
IEC60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Compliance
level
3 Vrms 150 kHz
to
80 MHz
ISM
outside
ISM
bands
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the system, including cables, than the
recommended separation distance calculated from the equation
appropriate for the frequency of the transmitter. Recommended
separation distances:
d
24
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of our product requires continued
operation during power mains
interruptions, it is recommended
that our product be powered from
an uninterruptible power supply or
a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
environment.
=
1
2 .
P
hospital