Herunterladen Inhalt Inhalt Diese Seite drucken

Stryker Flosteady Handbuch Seite 42

Spülpumpe für die arthroskopie
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Annual Inspection
EN
Fig. 12-1 Setup of basic function test
36
The layout of the test and its setup are depicted in Fig. 12-1.
1. Switch device on. Wait until the device check has finished.
2. Insert the cassette.
3. Suspend fluid bag at a height illustrated in Fig. 12-1 and connect bag with ir-
rigation tube.
4. Immerse the end with the Luer lock into the measuring cup.
5. Select a joint.
6. Set the nominal values:
• Nominal pressure = 150 mm Hg
• Max. flow = 1 l/min
7. Press the START/STOP key. Rotating arrow starts to turn.
8. Wait until the tube set is filled and a red dot appears in the rotating arrow.
9. Clamp off the instrument tube end in the measuring cup without stopping
the pump.
10. Empty the measuring cup.
11. Release the tube end. The pump is running.
12. Press the START/STOP key after 1 min. The measuring cup then contains ap-
prox. 1 l (±10%) of fluid.
The basic function test has been successfully completed once these values have
been reached.