Operating the Device
EN
Fig. 5-5
Hanging fluid bags
Fluid bags
(1)
Tap spikes
(2)
Tube clamps
(3)
Insert
Insert
35
35
Start Pumpe
Start Pumpe
tube set
tube set
mm Hg
mm Hg
mm Hg
mm Hg
18
5.5
Non-sterile tasks:
The tap spikes (see Fig. 5-5 "Hanging fluid bags", (2)) are used to connect the
tube sections with the fluid bags (1) (optional safe lock connection).
WARNING!
The device is only intended for use with flexible fluid containers. Do not use glass
containers as they might break. Fluid cannot flow quickly enough due to the vac-
uum being generated inside of the bottle. Risk of implosion.
WARNING!
Select an irrigation fluid suitable for the subsequent medical procedure.
The fluid bags are to be attached at a height of h 0.7 to 1.0 m (Fig. 5-5 "Hanging
fluid bags") (pump inflow at center of fluid fill level in container).
(1)
(2)
(3)
1. The inflow tube can take irrigation fluid from two fluid bags (1). Close both
clamps (3) at the branches of the inflow tube.
2. Grasp the tap spike (2) at the provided handle when connecting or discon-
1.5
1.5
necting.
150
150
3. Comply with sterile precautions when inserting the spike into a bag (optional
125
125
safe lock connection).
l/min
l/min
100
100
75
75
4. Open at least one clamp (3) at the irrigation tube.
50
50
25
25
5.6
0
0
Turbo
Turbo
Wash
Wash
If the tube cassette was not inserted, the function display depicts the message In-
sert tube set.
The device switches off automatically if the battery is used to operate the device

Connecting the Fluid Bags

Starting the Pump Function of the Device