Herunterladen Inhalt Inhalt Diese Seite drucken

Pump Connection; Interlock - Varian 929-0200 Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

12 Vdc

Pump Connection

The pump is connected to the controller rear panel via
the high voltage connector by a coaxial high voltage
cable assembly.
!
High voltage can cause severe injury or death. Before
installing or removing the connector:
Turn the power off.
When connecting the cable to 2 l/s pumps, or
larger, be sure that the ground spring is in place on
the high voltage connection of the pump.
Install the pump connector and secure it with
screws.
Plug the controller connector.

Interlock

The interlock is a safety device designed by Varian to
prevent serious damages to the operator due to high
voltage on the pump power connector.
Therefore, two serial electrically connected interlocks
are planned; both of them should be closed, so that the
pump can activate.
H.V. Cable Interlock consists of a Jack (A)
grounded through a cable connected to the metal
frame.
Remote ON/OFF Interlock consists of a P001
connector (included) whose 3 and 7 pins are short-
circuited (please refer to the previous figure).
In order to activate the pump, you should:
1. Plug Jack (A) into plug (B).
2. Plug P001 connector into J001 connector.
3. Place the main switch to ON.
J001
1
6
2
1 K
7
K
3
8
4
1 K
9
5
MicroVac Controller
P001 mating connector and J001 accessory connector
WARNING!
P001
1
6
2
7
3
8
4
9
5
J001
Interlock
Jack (A)
Interlock Jack position
22
TECHNICAL INFORMATION
GND
10V=1mA REC. LOG. OUTPUT
INTERLOCK
GND
1V=1KV ANALOG OUTPUT
H.V. ON SIGNAL
H.V. ON SIGNAL
B
87-900-089-01 (F)