CONTROLLER SPECIFICATIONS

Operation
Input:
Voltage
Frequency
Power
Output:
Voltage
Current
Power
Current Recorder
Output:
- Log Rec Out
(see attached
graphs)
Voltage Recorder
Output:
- Rec Out
-6
1x10
Mini, 2 l/s pumps
100, 120, 220, 240
Vac ±10%, 1-phase
47 to 63 Hz
40 VA maximum
3.300 Vdc nominal ±10%,
1 mA (short circuit)
1.2 W maximum
0 to 10 Vdc (10 V
correspond to 1 mA;
10 nA resolution
0 to 3.3 V corresponding to 0
to 3.3 KV with 1 V
corresponding to 1 KV in
linear mode)
Available at pins 4 and 8 of
rear panel Accessory
connector.
-5
1x10
Pump current/voltage diagram
Operating
temperature
Storage temperature
Fuse (mains)
Radio interference
suppression
Safety
Installation category
Pollution degree
Output Connector:
- High voltage
- Accessory
Cable
Weight:
-4
1x10
CURRENT (A)
18
TECHNICAL INFORMATION
0 °C to +45 °C
-20 °C to +70 °C
2
x
T
1A
(slow
disregarding the mains
EN 55011 class A group 1
EN 61000-4 -2/3/4
EN 61010 - 1
II
2
KINGS type:
P/N 10664-I for Model
929-0200 and 929-0201
J 101 input/output signal (9-
pin "D" type socket)
Mains, 3 meters long
1.7 Kg (4 lbs)
-3
1x10
1x10
87-900-089-01 (F)
blow)
-2