Problems
The irrigation pump goes on
and off erratically causing
abnormal pressure fluctuations.
There is too much water in the
pressure chamber.
There is insufficient pressure
in the system.
Cause
Water may have entered the
pressure sensing line.
An air leak in the line may
have developed.
An irrigation problem exists.
0
Possible solutions
Change the �One Day Set™�
ange the �One Day Set™�
tubing if water has entered the
pressure sensing line.
Check that the luer-lock on the
irrigation and pressure sensing
tube are closed. If necessary,
change the irrigation tubing
and start again.
Check the stop-cock on the
arthroscope, and the clamps
under the saline bags and
under the pressure chamber.