.. Pressure cycle mode.
This function increases the pressure by 60% for a two minute cycle. The pressure then returns
automatically to its original setting. The pressure cycle mode is activated when bleeding occurs.
The pressure cycle mode is
activated either by pressing the
�Pressure� pad...
.. Maintenance: cleaning & decontamination
Cleaning the fms solo
:
®
At the end of the operating day, switch the fms
solo
off, unplug the power cord and discard all
®
tubing. Wipe off the instrument's sides, front panel
and covers with a soft cloth that has been slightly
moistened with a neutral - pH detergent. Wipe again
using distilled water to rinse. DO NOT SOAK.
Do not use flammable agents when cleaning and
disinfecting the fms solo
.
®
... or using the remote control,
which is connected on the front
panel of the pump...
The remote control is not
autoclavable and not soakable.
A sterile disposable remote
control cover must be used (see
FMS Ref. 1087)
Foot pedal: cleaning and decontamination
After each operation, clean the foot pedal and the
power cord with a neutral-pH. The foot pedal can be
immerged, except the electrical connections which
must remain dry.
DO NOT AUTOCLAVE.
Do not twist the cable when storing.