.. Recommended initial settings.
Joint
Shoulder Joint
Acromioplasty
Knee Joint
Hemarthrosis
Wrist
Elbow, Ankle
Hip

.. Pressure alarm

High pressure alarm.
If the high pressure safety level
of 140 is exceeded, the pump
stops and the red ALARM (6)
LED lights up. The operator
should lower the pressure to let
the pump automatically restart.
If the pump does not restart
when the pressure drops below
140, restart the pump manually
by turning the pump ON and
OFF again with the POWER pad.
Note: numbers in ( ), refer to diagram on page 12.
Pressure level
with Tourniquet
30
50–60
30
40
Low pressure alarm.
If the pressure falls below the
low pressure safety level of 10,
the red ALARM LED lights up,
<Pres L> flashes on the LCD
display, and the pump stops.
Check that the luer-lock and
the filter near the pressure
transducer are in proper
working order.
If both are satisfactory press
Fill Chamber to stop the alarm
and Run/Stop pad to restart
the unit.
2
Pressure level
without Tourniquet
60
60
65
50–60
65
65
65
Caution:
If water has accidentally
entered the tubing leading
to the pressure transducer,
change the "One Day Set™"
and restart the system.