Pro-care 2 / English
NOTE 1: U
is the a.c. mains voltage prior to the application of the test level
T
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by ab-
sorption and reflection from structures, objects and people
a)Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoret-
ically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an elec-
tromagnetic site survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be observed to ver-
ify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the device.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
15
Where P is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufac-
turer and d is the recom-
mended separation distance
b
in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
a
survey ,
should be less than
the compliance level in each
frequency ranged.
Interference may occur in the
vicinity of equipment marked
with the following symbol: