The actions carried out during initialization are displayed in clear text in the
measured value display and the initialization progress is shown by a moving
bar.
The scale must not be loaded when it is being switched on.
The device is switched off by disconnecting the supply voltage from DWS2103.

7.2 Configuration of the load cells

During initial commissioning, all AED/FITs have the same address (31), a
measured value display is not possible and the display shows an initialization
error [INIT ERROR].
Each load cell must be assigned its own address. This is possible with the aid
of the load cell serial number.
Proceed as follows:
• The parameter menu must be called via the hidden key
• "DEVICE SELECTION" parameter menu
If the load cells have already been assigned individual addresses (e.g. with
the AED panel), measured values are displayed after switching on and you
can start directly with scale adjustment.

7.3 Zero setting the scale

Press the F4 key until the symbol >0< appears above the F3 key.
Press the F3 key to set the measurement value to zero.
The displayed gross value is zero immediately after zero setting.
Range of zero setting, lower limit
MODE ³
LEGAL-FOR-TRADE
NO (not leg. f. trade)
− 20 %
OIML, NTEP
− 2 %
The % figures relate to the nominal weighing range (parameter "SCALE PA-
RAMETERS → PARAMETER → NOMINAL VALUE")
Execution is dependent on standstill recognition (Chapter 7.9, page 26 ).
The net display is deactivated.
NOTE
Zero setting is only possible when the function in the "PARAMETER MENU →
FUNCTION KEYS → ZERO SETTING" is set to ACTIVE.
HBM
NOTE
Range of zero setting, upper limit
+ 20 %
+ 2 %
DWS2103
A2936-1.0 en/de