Error Codes
Error codes are displayed dynamically, for 2 seconds, or permanently.
After a code is displayed dynamically or for 2 seconds, the program returns
automatically to the normal operating mode.
Display
Cause
H
The load exceeds the weighing
L or Err54 Load is below the weighing
range
Err 01
Data output not compatible
with output format
Err 02
Calibration/adjustment
condition not met, e.g.,
– not tared
– the pan support is loaded
Err 03
Calibration/adjustment could
not be completed within a
certain time
Err 30
Interface port for printer
output is blocked
Err 31
Peripheral device not
responding (interface hand-
shake interrupted; XOFF, CTS)
Err 50
Temperature compensation
overflow/underflow
Err 53
Temperature compensation
not functioning
Err 55
Output from weighing ADC
too high
Err 79
Dryer adjustment data not
found
Err 241,
Weighing system parameters
Err 243
(EEPROM) defective
Err 2xx
Internal error
Err 340
Operating parameters
(EEPROM) incorrect
Err 342
Operating parameters
(EEPROM) incorrect except
adjustment parameters
If any other errors occur, contact your local Denver Service Center.
Solution
Unload the pan support capacity
Place the pan support on the
weighing system
Change the configuration in the
Setup menu
Calibrate only when zero is
displayed
Select TAR to tare
Unload the moisture analyzer
Allow the scale to warm up
again and repeat the adjustment
Have the port setting changed
by Denver Customer Service
Send XON, release CTS
Contact your local Denver
Service Center
Contact your local Denver
Service Center
Contact your local Denver
Service Center
Contact your local Denver
Service Center
Switch the analyzer off and then
on again. If the error persists,
contact your local Denver
Service Center
Contact your local Denver
Service Center
Contact your local Denver
Service Center
Contact your local Denver
Service Center
29