Operation
Example: Analysis with Specified Drying Time
The specified drying time in the example is 10 minutes.
Final temperature:
Start of analysis:
End of analysis:
Display mode:
Step
1. Switch on the moisture
analyzer
2. Set the drying parameters
(see "Setting the Drying
Parameters" in the chapter
entitled "Configuration")
3. Open the sample chamber
and place an unused sample
pan on the pan support
4. Tare the sample pan:
select TAR
...and confirm
5. Distribute approx.
2 g sample evenly on
the sample pan
6. Close the sample chamber
The printout header is
printed: see next page
22
130°C
Without stability after the cover is closed
After 10 minutes
Moisture