Features
Temperature during heating
– Adjusted to defined specified
temperature during the analysis
process
Start of analysis
– With stability after the u key
is pressed:
When START is shown in the
function line and you press u
to confirm, the initial weight is
stored at stability regardless of
whether the cover is open or
closed. Measurement begins as
soon as the cover is closed.
– Without stability after the cover
is closed:
A symbol shown in the graphic
symbol display prompts you to
close the cover once the initial
weight condition is met. The initial
weight is stored without stability
as soon as the sample chamber is
closed, and analysis begins.
End of Analysis with Shutoff
Parameters
– Fully automatic mode
– Timer mode
Fully automatic mode:
Use the fully automatic mode when
loss of weight on drying follows a
clearly delineated curve which can
be unambiguously evaluated (see
below).
Weight
Moisture (evaporation)
Automatic shutoff
Timer mode:
The analysis ends as soon as the
specified time has elapsed.
Display Mode for Result
The following units can be selected
for displaying analysis results:
– Moisture
– Dry weight
– Ratio
– Residual weight
Print Intermediate Results
Intermediate results can be printed
either at user-definable intervals or
by pressing the r key.
Time
%M
%S
%MS
g
19