Herunterladen Inhalt Inhalt Diese Seite drucken

Denver Instrument IR-35 Betriebsanleitung Seite 26

Elektronischer feuchtebestimmer
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Step
The prompt for calibration
weight is displayed
8. Open the hinged cover
9. Place the 30-weight on
the weighing system
Minus sign –:
weight value too low
Plus sign +:
weight value too high
No sign: weight value OK
The weight unit symbol
(g) is displayed at the
end of adjustment
Printout after calibration
and adjustment
*) Date and time included
only if a Denver printer
is used
10. Unload the analyzer
Close the cover
11. Quit calibration/adjustment
26
Key (or instruction) Display
c
o
P
˚
C
min
%
g
/l
CAL SET PRG
END
o
P
˚
C
min
%
g
/l
CAL SET PRG
END
o
P
˚
C
min
%
g
/l
CAL
SET PRG
END
--------------------
23.08.2005
Model IR35M-000230V1
Ser. no.
99992581
Ver. no.
00-33-01
ID
--------------------
External calibration
W-ID
Nom.
+
30.000 g
Diff. +
0.001 g
External adjustment
Diff. +
0.000 g
completed
--------------------
Name:
--------------------
A
min
E
MS
START
TAR
A
min
E
MS
START
TAR
A
min
E
MS
START
TAR
*)
10:51

Quicklinks ausblenden:

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis