Adjusting the Analyzer

Heating Element Adjustment

The procedure for adjusting the
heating element is described in
the instructions supplied with the
temperature adjustment set.
24

Weighing System Adjustment

To adjust the weighing system,
perform calibration and adjustment
as described in the following.
Purpose
Calibration is the determination of
the difference between the weight
readout and the true weight (mass)
of a sample. Calibration does not
entail making any changes within
the weighing system.
Adjustment is the correction of the
difference between the measured
value displayed and the true weight
(mass) of a sample, or the reduction
of the difference to a level within
specified permissible error limits.
Features
Calibration is performed externally
with the following weight value:
– IR-35: 30 g
You can have calibration and
adjustment results documented as
a ISO/GLP-compliant printout (see
the page after next for an example).
External Calibration and Adjustment
with a Factory-Defined Weight
Externally calibrate and adjust the
weighing system using a 30-g
calibration weight.