8. Alarm Display

Display
(Lights on)
(Flashing) Error in stored data
22
(E)
Trouble
Measuring unit not
connected
Excess speed
Overflow
Power failure
Error in reference point
detection
Causes/Remedy
The measuring unit is not connected.
Turn off the power, connect the measuring unit,
and then turn on the power again. The display
value is reset to zero.
The maximum response speed is exceeded at
the measuring unit side.
Perform resetting operation.
(The same condition may occur when the
machine is subjected to a major shock.)
When the display has overflowed, an "F" is
added to the highest digit.
Use in a range where an "F" is not added.
The power fails momentarily during
measurement.
Perform resetting operation.
The stored data has been changed by noise or
other cause. Redo the settings starting from the
basic settings.
If this error is displayed frequently, the memory
may be damaged. Contact your vendor.
: Error code (1 to 9, A to F)
This is displayed when a measuring unit without
a reference point is connected or when the
reference point signal wire in a measuring unit
with a reference point is broken. Connect a
measuring unit with a reference point. If this
does not correct the problem, contact your
vendor.