6. Specifications
Function
General-
Input connector
purpose
input
General-
Output connector
purpose
output
Linear compensation
Scaling
Key lock
Current value store
Display at power ON
Flicker control
Expansion units
Power save
Power supply
Power consumption
Operating temperature range
Storage temperature range
Mass
20
(E)
Description
Phoenix Contact terminal block connector, 9 pins (Including external reset
and external preset value call (preset recall))
The function can be selected for inputs 1 to 3.
Input 1 : (for axis A) Hold function (latch, pause), Restart, Display mode
switching, External reference point load
Input 2 : (for axis B) Hold function (latch, pause), Restart, Display mode
switching, External reference point load
Input 3 : (for all axes) Hold function (latch, pause), Restart, Display mode
switching
Phoenix Contact terminal block connector, 5 pins
The function can be selected for outputs 1 to 4.
Outputs 1 and 2 : (for axis A) Alarm, Display mode, Reference point
detected signal, Reference point alarm, Signal when
going past zero point
Outputs 3 and 4 : (for axis B) Alarm, Display mode, Reference point
detected signal, Reference point alarm, Signal when
going past zero point
A fixed compensation amount is applied to the measuring unit's count
value.
Compensation amount Standard: ±600 µm/m (Expanded: ±1000 µm/m)
Scaling factor: 0.100000 to 9.999999
It is possible to set and cancel the key lock.
It is possible to set whether to store the current value at power OFF.
display or count display can be selected.
When the minimum digit of the display value is unstable, the average value is
displayed.
BCD, Comparator
The display is turned off when no operations are made for a preset time.
(The time can be set.)
DC 12 V Rating 0.75 A Max. 1 A
AC 100 V - 240 V ±10 % (When using the AC adaptor (option))
Max. 32 VA (connected to AC power supply)
0 to 40 °C (no condensation)
–20 to 60 °C (no condensation)
Approx. 1.5 kg