Surgical handpiece
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
( 3 ) Sterilization
Autoclave sterilization is recommended.
Sterilization is required first time you use and after each patient as noted below.
This handpiece can be cleaned and disinfected with a Thermo-Disinfector.
This handpiece can be autoclavable up to Max.135°C.
1 ) The dirt of the handpiece surface is discarded with a brush ( metal is improper ). Clean handpiece using water or Isopropyl
alcohol.
2 ) Lubricate the handpiece. ( Refer ( 2 ) Lubrication )
3 ) Insert into an autoclave pouch. Seal the pouch ( Only if using a vacuum autoclave )
4 ) Autoclavable up to max.135°
ex. ) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5 ) Keep the handpiece in the autoclave pouch to keep clean before use.
14