Herunterladen Inhalt Inhalt Diese Seite drucken

Correct Use - IKA RV 10 series Betriebsanleitung

Rotationsverdampfer
Vorschau ausblenden Andere Handbücher für RV 10 series:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
Safety lift
When the device is switched off or the power supply disconnected,
the internal safety lift removes the evaporating flask from the
heating bath.
The safety lift at loss of power is designed for a maximum total
weight (glassware and solvent) of 3.1 kg.
Sample calculation of the maximum load with vertical glassware
and a 1 litre flask:
Condenser + receiving flask + evaporating flask + fittings=
1200 gr + 400 gr + 280 gr + 100 gr = 1980 gr
Maximum loading of solvent = 3100 gr – 1980 gr = 1120 gr
Due to the design, operation of the safety lift cannot be
guaranteed for higher loads!
When using other types of condensers such as dry ice or intensive
condensers, and also when using return distillation distributors
with slip-on condensers, it may be necessary to reduce the load by
the amount of the added weight of the glass apparatus.
Thus, prior to distillation, check whether the lift goes up without
power when laden with the glassware and distillation material.
The safety lift must be checked daily prior to operation. Use
the motor to move the lift to the lowest position and press the
"Power" key on the front plate or the main switch on the back
right side of the device using the maximum total weight of 3.1 kg.
 The evaporating flask is lifted out of the heating bath.
When using for the first time or after a long period of inactivity,
proceed as follows: use the motor to move the lift several times
to the top and bottom positions by pressing the lift function keys
prior to interrupting the power supply.
If the safety lift does not work, please contact the IKA
department.
For the evaporating equipment (evaporating flask plus contents),
the maximum permissible weight is 3.0 kg! Loads greater than this
risk breakage of glass at the vapor tube!
Ensure that the safety lift has been powered down.
When working with large loads, always use low speeds.
Unbalanced loads can result in breakage of the vapor tube!
• Use
Together with the accessories recommended by IKA
device is suitable for:
- quick and gentle distillation of liquids,
- evaporation of solutions and suspensions,
- crystallization, synthesis or cleaning of fine chemicals,
- drying of powder and granulate material,
- recycling of solvents.
Mode of operation: Tabletop device
• Range of use
- Laboratories
- Pharmacies
®
Service

Correct use

- Schools
- Universities
• A vacuum may be formed inside the glassware in the case of
power outage. The glassware must be vented manually.
CAUTION
• Set the speed of the drive so no tempering medium is sprayed
out as a result of the evaporation flask rotating in the heating
bath. If necessary reduce the speed.
• Do not touch rotating parts during operation.
• Imbalance may result in uncontrolled resonance behavior of
the device or assembly. Glass apparatus may be damaged or
destroyed. In the event of unbalance or unusual noises, switch
off the appliance immediately or reduce the speed.
• The appliance does not start up again automatically following a
cut in the power supply.
• The device is only disconnected from the power supply network
if the device power switch is off or the plug is pulled out.
• The socket for the mains cord must be easily accessible.
• Always check evaporation processes before starting automatic
operation. Automatic operation must not be used for unfamiliar
evaporation processes. The following automatic modes are
available in addition to manual mode:
- 100% distillation,
- Volume-/quantity-dependent distillation.
For protection of the equipment
• The voltage stated on the type plate must correspond to the
mains voltage.
• Socket must be earthed (protective ground contact).
• Removable parts must be refitted to the appliance to prevent
the infiltration of foreign objects, liquids, etc.
• Protect the appliance and accessories from bumps and impacts.
• The appliance may only be opened by experts.
This device is suitable for use in all areas except:
, the
®
- Residential areas
- Areas that are connected directly to a low-voltage supply net-
work that also supplies residential areas.
The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not supplied or
recommended by the manufacturer;
- If the device is operated improperly or contrary to the manufac-
turer's specifications;
- If the device or the printed circuit board are modified by the third
parties.
Never operate the device when the
evaporation flask is rotating and the lift
is raised. Always lower the evaporation
flask into the heating bath first before
starting the rotation drive. Otherwise
hot tempering medium may be sprayed
out!
34

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Rv 10 controlRv 10 auto

Inhaltsverzeichnis