CLEANING AND DISINFECTION

DANGER
Danger of electrocution
Liquids can conduct electricity, thereby posing a risk of elec-
tric shock.
• Always switch the compressor off and disconnect the
power plug from the mains socket before starting to clean
it.
If the inhalation system is used by only one patient, follow the
instructions for cleaning and disinfecting in the section "Single
patient use" [see: Single patient use, page 27].
If the inhalation system is used by several patients, follow the
instructions for cleaning and disinfecting in the section "Multiple
patient use" [see: Multiple patient use, page 32].

4.1 Single patient use

Cleaning and disinfection cycles
Compressor housing
Nebuliser (without con-
nection tubing) and mask
Air filter
– 27 –
Clean when visibly dirty
– Clean immediately after every use
– Disinfect at least once a day
Replace after 200 operating hours
(approx. 1 year)