Type 8619
Assembly instructions
Montageblatt
1410/0_EU-ML_00566607
assembly kit 00564596 — assembly instructions of the
kit and set up instructions of the 8619
NOTE
Risk of material damage if the cable glands of the 8619 are loose. The body of the cable glands are screwed into the
housing at the factory with a torque of 5.5 Nm
• Before installing the housing on the pipe, check that the bodies of the cable glands are tightened. If the bodies of the cable
glands are loose, tighten them with a torque of 5.5 Nm ± 20 %.
The assembly kit allows you to install the 8619, wall-mounting versions, on horizontal or vertical pipes whith an external dia-
meter comprised between 30 mm and 110 mm.
→
Choose a location such that:
• The surface temperature of the pipe remains below 100°C.
• The display is at eye level.
• There is sufficient space to open the housing by 180°.
1
fastening plate
90 mm
135 mm
3
2
fastening brackets
90 mm
135 mm
Step 1: Remove the fastening plate from the device.
→
Press the tab to unlock the device.
→
Lift the device.
→
Separate the device from the fastening plate.
Step 2 for an horizontal pipe: Mount the 2 assembly
kits.
→
Insert the 4 screws in the holes drilled in the fastening
plate.
→
Insert each fastening bracket in 2 screws.
→
Respect the 90 mm distance between two screws,
as shown in the drawing on the left.
→
Insert a locking-washer and a nut on each screw.
→
Tighten the 4 screws with a maximum torque of
5.3 Nm.
Step 2 for a vertical pipe: Mount the 2 assembly kits.
→
Insert the 4 screws in the holes drilled in the fastening
plate.
→
Insert each fastening bracket in 2 screws.
→
Respect the 135 mm distance between two screws,
as shown in the drawing on the left.
→
Insert a locking-washer and a nut on each screw.
→
Tighten the 4 screws with a maximum torque of
5.3 Nm.