SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
WARNING
Use this cylinder in accordance with the HomeFill
Owner's Manual. Open valve slowly, Close valve after
each use and when empty. Secure cylinder during storage
and use. No smoking in cylinder area. Keep away from
heat, flame or spark. Keep out of the reach of children.
DO NOT drop.
Transfilling of this gas is performed by HomeFill only.
Cylinders that have been refinished and/or exposed to
elevated temperatures MUST be hydrostatically tested
before filling. Cylinder temperature should not exceed
52° C (125° F). Aluminum cylinders subject to the action
of fire or heated temperatures in excess of 177° C (350° F)
MUST be withdrawn from service and condemned by
trained personnel. DO NOT use caustic paint strippers.
Valve and safety relief devices MUST be removed and
replaced by trained and authorized personnel. DO NOT
alter or change valve assembly.
All cylinders MUST be inspected before attempting to fill;
otherwise, injury or damage may occur.
NEVER smoke in an area where oxygen is being
administered.
NEVER use near any type of flame or flammable/explosive
substances, vapours or atmosphere.
NOTE: For this procedure, refer to FIGURE 4.1.
The Hydrostatic Testing Date
Aluminum cylinders MUST undergo testing every five years.
HomeFill
12
Part No 1141477