6.4.2
Sterilisation
After cleaning/disinfection, pack the dismantled nebuliser in sterilisation
packaging (disposable sterilisation packaging, e.g., foil/paper sterilisation
bags). The sterilisation packaging must conform to DIN EN ISO 11607 and
be suitable for steam sterilisation. Then sterilise the nebuliser using the fol-
lowing sterilisation procedure:
Recommended procedure: steam sterilisation
Sterilisation temperature: 121°C (to be maintained for at least 20min.) or
This procedure has been validated according to DIN EN ISO 17665 as well
as DGKH guidelines for the validation and routine monitoring of sterilisation
processes using damp heat for medical devices.
Storage
Keep the sterilised nebuliser in a dry, dust-free place protected from contam-
ination.
6.4.3
Care of the connection tubing
• In the case of visible condensation in the tubing, pull it off the nebuliser
after treatment and switch on your PARI compressor.
• Leave the compressor running until any condensation in the tubing has
been removed by the air-flow.
• Replace the connection tubing in the case of a change of patients or if it
becomes contaminated (part no. 041G4550, see Chapter 9).
132°C/134°C (to be maintained for at least 3min.)
max. 137°C.