2. Keep the brake button pressed and move the head
slide until a length value appears in the display.
ATTENTION!
Incorrect measurement as a result of
incomplete calibration.
1.
The height value displayed does not
correspond to the actual position of the head
slide.
− Complete the calibration as described in this
section.
3. Position the calibration rod supplied (length 0.8 m)
vertically on the foot mat and hold it in this position.
4. Keep the brake button pressed and push the head
slide onto the calibration rod.
5. Simultaneously press the Enter key (send/print)
and the arrow key (hold/zero) to open the menu.
The last menu item selected is shown in the display
(here: "bEEP").
6. Press the arrow key (hold/zero) until "CAL"
appears in the display.
7. Confirm your selection with the Enter key (send/
print).
8. Use the arrow key (hold/zero) to select the "Yes"
setting.
9. Confirm your selection with the Enter key (send/
print).
The display "Auto" appears.
10. Confirm the display using the Enter key (send/
print).
The device is calibrated. You can perform height
measurements with the device.
Before you get started ... • 57