Herunterladen Inhalt Inhalt Diese Seite drucken

ABB TB82PH Bedienungsanleitung Seite 12

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

TB82PH ATEX Product Instruction Manual
AP100, AP200, AP300 7660 and 7650 sensors have different color-coding in the cable assemblies than TB5 and TBX5
sensors. Use the table below to properly connect the sensor cable to the TB82PH.
TB82PH Terminal TB2
Color
Number
Description
Blue
1
Measuring electrode
Shield for measuring
Yellow
2
Black
3
Solution (diagnostic)
Green
4
2
Red
5
2
White
6
Heavy Green
7
Shield (Screen)
None
8
Notes:
1. BNC adapter required on sensors provided with BNC termination (See Figure 9).
2. Red and White wires not present on sensors sold without temperature compensators
3. Cut away and discard gray wire on AP100 sensors (if present)
4. Place both red wires in terminal 5
Note.
TB5, AP100, AP300 and 7650/7660 sensors do not have a
solution ground rod required to fully utilize all the TB82PH
on-line sensor diagnostics.
transmitter for use with these sensors the DIAGS
(diagnostics) choice must be set to OFF.
Configure Mode section for details.
Note.
If an automatic temperature compensator is not used the
TB82PH must be configured for manual temperature
compensation. See the Configure Mode section for details.
12
ABB Models
ABB Models
TB5
1
Blue
None
electrode
1
Reference
Black
None
Ground
RTD
Red
RTD
White
None
Not Used
None
When configuring the
See the
ABB Models
ABB Models
3
TBX5
AP100
Blue
Clear
Yellow
None
Black
Black
Green
None
Red
Red
White
White
Heavy Green
None
(Shield)
None
None
ABB Models
ABB Models
AP200
AP300
Clear
Blue
None
None
Red
Black
Green/Yellow
None
Gray & Green
Red
White
White
None
None
None
None
TB82PH-ATEX-EN-B
ABB
7650/7660
Clear
None
Black
None
4
Red
(2 total)
White
Yellow
None

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis