Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 26
CAUTION
NOTICE
10
Changing the batteries
CAUTION
1) Turn OFF the power.
2) Loosen the battery cover screw on the bottom of the control unit then remove the
battery cover.
3) Remove old batteries.
4) Insert new batteries, following the positive (+)/negative (-) indication in the bat-
tery box.
CAUTION
5) Attach the battery cover to the control unit and tighten the battery cover screw.
14
is in a pouch. This is to prevent possible discoloration and damage
to the product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature,
humidity, ventilation, and sunlight. The air should be free from
dust, salt and sulphur.
Do not heat or cool the product too quickly. Rapid change in tem-
perature could cause damage to the product.
If the sterilizer chamber temperature may exceed 135°C during the
dry cycle then delete the dry cycle.
Autoclave sterilization is recommended for the product. The validity
of other sterilization methods is not confirmed.
Do not touch the product immediately after autoclaving as it will be
very hot and must remain in a sterile condition.
NSK recommends Class B sterilizers as stated in EN13060.
When the Battery indicator flashes, immediately replace original
batteries with new dry cell batteries.
Make certain the positive (+) and negative (-) terminals are cor-
rectly aligned.
When it is hard to insert, do not forcibly insert the battery as it may
be inserted incorrectly.

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis