Fault detected
Device does not operate
(LCD screen off)
No spray
No air
Power fault /
weak vibrations
16
Possible causes
Power cord not plugged in properly.
Power switch in position O.
Power receptacle fuse(s) blown.
Internal fuse blown.
Irrigation ON/OFF button activated.
air function selected.
Irrigation solution tanks empty.
Faulty pump.
Blocked irrigation circuit.
Blocked tip or file.
Wrong choice of tip.
Handpiece cord blocked.
Faulty air connection.
Irrigation ON/OFF button deactivated.
air function activated.
Blocked irrigation circuit.
Blocked tip or file.
Wrong choice of tip.
Handpiece cord blocked.
Tip worn or distorted.
Incorrect power adjustment.
Incorrect use: incorrect angle of attack or
inadequate pressure on the tooth.
Liquid or moisture present between
handpiece and cord.
Solutions
- Check the power socket.
- Return to SATELEC after-sales.
Place the power switch in position I.
Replace the power receptacle fuses.
Return to SATELEC after-sales.
Press the Irrigation ON/OFF button.
Press the air button.
Fill the tanks with irrigation solution.
Return to SATELEC after-sales.
Return to SATELEC after-sales.
Unblock the tip or file.
Check the tip.
Replace the cord.
Check the air connection.
Press the Irrigation ON/OFF button.
Press the air button.
Return to SATELEC after-sales.
Unblock the tip or file.
Change the tip.
Replace the cord.
Replace the tip.
Refer to the TipBook for the necessary
information.
Refer to the TipBook for the necessary
information.
Dry the electrical contacts thoroughly.