Herunterladen Inhalt Inhalt Diese Seite drucken

Downloading Data - Elcometer 415 Gebrauchsanleitung

Farb- und pulverbeschichtungsmessgerät (zum messen auf glatten oberflächen)
Vorschau ausblenden Andere Handbücher für 415:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 34
7 CALIBRATING THE GAUGE (continued)
7.4 CHECKING THE GAUGE CALIBRATION
1
Place a calibration foil on to the steel or aluminium check piece
supplied.
2
Whilst holding the gauge perpendicular, gently place the probe
onto the foil.
3
The gauge should read the value printed on the foil label ±1-3%
or ±2.5
μm (±0.1mil)
If the gauge does not display a valid reading, it may require
re-calibration - see Sections 7.2 & 7.3.

8 DOWNLOADING DATA

8.1 USING ELCOMASTER ON A PC
Using ElcoMaster - supplied with each gauge and available as a free
download at elcometer.com - the Elcometer 415 can simultaneously
transmit the live reading via USB or Bluetooth (Model T only), to a PC
for report generation. For more information on ElcoMaster visit
www.elcometer.com.
8.2 USING ELCOMASTER MOBILE APPS - MODEL T
Ideal when out in the field or on-site, using the ElcoMaster Mobile App
users can:
Store live readings directly on to a mobile device and save them
into batches together with GPS coordinates.
Add photographs of the test surface
Map readings on to a photograph or diagram
'Collect Design' function in ElcoMaster - see Appendix 2 'ElcoMaster
Collect Design' on page en-13)
Inspection data can be transferred from mobile to PC for further
analysis and reporting.
For more information on ElcoMaster Mobile Apps visit
www.elcometer.com
www.elcometer.com
- whichever is the greater.
®
®
®
®
(via the mobile device's camera).
(when using the
®
®
Compatible with smart phones
and tablets running Android 2.1
or above. To install, download
via www.elcometer.com or using
the Google Play™ Store app,
and follow the on screen
instructions.
R
®
®
®
en-8

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis