5. Cleaning
Warning
24
l Use by more than one mother without
adequate reprocessing may be a health
risk and could cause cross-contamination.
l If there is no milk, condensation or visible
residues in the tubing, there is no need
for automated cleaning and disinfection
or sterilisation. Medela recommends that
the tubing is disposed of and replaced
after max. 6 months.
If there is milk, condensation or visible
residues in the tubing, dispose of and
replace.
Caution
l Only use drinking-quality water for
cleaning.
Note
l Only sterilise clean and intact parts. Do
not stack parts on top of one another.
To avoid damage, do not apply external
pressure.
5.8 Reusable pump set – Auto-
mated cleaning and disinfection
1
If not done yet, disassemble the re-
usable pump set into its individual parts
(bottle, breast shield, connector, valve
head, valve membrane, membrane
cap, protective membrane and tubing).
Gently remove the white membrane
from the yellow valve head.
Check all parts for damage.
Replace damaged or worn parts.