3.4. Emptying Flexel
3D 1000L Bags
Sartorius Stedim Biotech recommends the
dispensing through the bottom.
a) Take the bottom port tube and position it
vertically. Open the connector protective
outer-wrap using the reversible cable tie
on the dispensing line and carry out the
connection according to the standard
procedures that apply on the site
(See Fig. C).
b) Open the clamp (See Fig. A). Drain the
bag - Gravity can be used, but for faster
draining Sartorius Stedim Biotech recom-
mends the use of a peristaltic pump.
(See Fig. 17)
c) The bag collapses as it is drained. The
sloped bottom of the Palletank
position of the drain hole, provide nearly
100% fluid recovery. This efficiency can
be improved by lifting the bag at the end
of the operation. (See Fig. 18)
d) When the dispensing is complete, clamp
the line (See Fig. B) and disconnect the
main drain line from the dispensing tube
®
of the Flexel
3D Bag.
e) Open the bottom gates by removing the
retaining clip from the retaining clip fas-
tening point; then grasp the retaining
clip fastening point and swing the
bottom gates open (see Fig. 7).
f) Lift the collapsed bag and remove it from
®
the Palletank
(See Fig 18).
g) Discard the Flexel
®
3D Bag in accordance
with the procedures used on your site.

4. Troubleshooting Guide

The bag will not fill.
> Check the clamp is open and the tube is
not pinched.
When filling through the bottom, the
bag comes out of the bottom gate.
> The edge of the jaw is not closed inside
the groove of the bottom port. Position
it correctly in the groove.
The bag does not open out correctly
during filling:
The port system is incorrectly aligned
> Open the bottom gate, turn the bag to
bring the port system opposite and close
the bottom gate.
The bottom of the bag is incorrectly
positioned.
®
, and the
> Place the corners of the bag in the cor-
ners of the Palletank
only 5–6% of nominal volume is filled
before adjusting, otherwise bag failure
may occur).
Many folds appear in the bag during
filling.
> The bag is too big, use the bag suited to
the size of the Palletank
The bag does not come up against the
walls.
> The bag is too small, use the bag suited
to the size of the Palletank
The bag will not drain.
> Check that the clamp is open and that
the tube is not pinched.

5. Precautions of use

> To ensure trouble-free operation, the
Sartorius Stedim Biotech Flexel
must be used with the Sartorius Stedim
Biotech Palletank
> The Palletank
before inserting the Flexel
Traces of cleaning agents left on the
Palletank
Flexel
> It is under the user's responsibility to
verify the compatibility of the Flexel
Bag materials with the solutions used in
your process. Sartorius Stedim Biotech
can provide guidelines for chemical
compatibility, bio-compatibility and
extractables.
> Avoid leaving sharp objects (scissors,
®
(make sure that
cutters, scalpels, etc.) in the vicinity of
the Flexel
> The Flexel
by its connections (i.e. tubing).
> Avoid folding the Flexel
than the original folds. Do not fold the
®
bag in the direction of the "gussets"
.
(gusset = heat sealed seams).
> When positioning the Flexel
the Palletank
®
.
the initial fill, handle the bag gently and
do not force fit or pull hard.
> The Flexel
should not be re-used.
®
3D Bag
®
.
®
must be clean and dry
®
3D Bag.
®
can potentially damage the
®
3D Bag.
®
3D Bag.
®
3D Bag is not to be handled
®
3D Bag other
®
3D Bag in
®
and when adjusting it after
®
3D Bag is for single-use and
®
3D
5