the graphical classification on the device always shows
the higher category; for the example given this would be
‚High normal'.
Blood pressure
Systole
value category
(in mmHg)
Setting 3:
severe hyper-
≥ 180
tension
Setting 2:
moderate hyper-
160 – 179
tension
Setting 1:
140 – 159
mild hypertension
High normal
130 – 139
Normal
120 – 129
Optimal
< 120
Source: WHO, 1999 (World Health Organization)

6. Saving, retrieving and deleting results

• The results of each successful measurement are
stored together with date and time. With more than
60 items of measured data, the earliest items of data
measured are lost.
Diastole
Action
(in mmHg)
seek medical
≥ 110
attention
seek medical
100 – 109
attention
regular monito-
90 – 99
ring by doctor
regular monito-
85 – 89
ring by doctor
80 – 84
self-monitoring
< 80
self-monitoring
• In order to retrieve measurement results again, press
the memory key M. Initially, the average value of the
last three blood pressure readings is displayed (Dis-
play ). Press the memory key M again to display the
latest measurements with date, time and graphic WHO
classification and to hear the values announced. You
can adjust the volume during the announcement by
pressing the volume key + and -.
• You can clear the memory by pressing and holding +
and - simultaneously for 5 seconds after pressing the
memory key M and selecting the user.
• If you wish to change the user memory, see "Select
memory".

7. Error messages/troubleshooting

In case of faults, the
Error messages may appear if:
• systolic or diastolic pressure could not be measured
(
or
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ( or Lo appears on the display)
• the cuff is fastened too tightly or loosely ( Er3 or Er4
appears on the display)
• the pump pressure is higher than 300 mmHg ( Er5
appears on the display)
• pumping up takes longer than 160 seconds ( Er6
appears on the display)