• there is a system or device error ( ErA ,
appears on the display)
• the batteries are almost empty
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk.
Re-insert the batteries if necessary, or else replace them.
Technical alarm – description
Should the recorded blood pressure (systolic or diasto-
lic) lie outside the limits specified in the section "Techni-
cal specifications", the technical alarm will appear on the
display indicating either " " or " Lo ". In such cases, you
should seek medical assistance and check the accuracy
of your procedure.
The limit values for the technical alarm are factory set
and cannot be adjusted or deactivated. These alarm limit
values are accorded second priority under the standard
IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not
be reset. The signal shown on the display will disappear
automatically after about 8 seconds.

8. Cleaning and storing the instrument

• Clean your blood pressure monitor carefully using a
slightly damp cloth only.
• Do not use any detergents or solvents.
, Er7 or Er8
.
24
• Never hold the instrument under water as otherwise
liquid can penetrate and damage the instrument.
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
Measurement
method
Measurement range
Display accuracy
Measurement inac-
curacy
Memory
Dimensions
Weight
Cuff size
BC 19
Oscillometric, non-invasive blood
pressure measurement on the
wrist
Cuff pressure 0 – 300 mmHg,
systolic 60 – 260 mmHg,
diastolic 40 – 199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard devia-
tion according to clinical testing:
systolic 8 mmHg /diastolic
8 mmHg
2 x 60 memory spaces
L 85 mm x W 70 mm x H 31 mm
Approx. 140 g (without batteries)
14.0 to 19.5 mm