10 Technical data

10.1 Typical Product Parameters

Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
Oxygen Output Concentration Levels:
Maximum Outlet Pressure:
Low Flow Alarm:
Potential Obstruction Alert:
Power Consumption:
Pressure Relief Mechanism Operational
at:
1195656-A
230 VAC + 10, –15% (253 VAC/195.5 VAC), 50 Hz
1.5 A
40 dBA max
Up to 6,000 ft (1828 m) above sea level without degradation of concentration levels. Not recommended
for use above 1828 m (6,000 ft).
Atmospheric pressure range: 101.33 kPa — 81.22 kPa
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
Concentration levels achieved after initial warm-up period (approximately 30 minutes)
34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)
0.5 to 5 L/min (maximum). For flowrates less than 1 L/min, we recommend the use of the Invacare
Pediatric Flowmeter Accessory (IRCPF16AW).
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a potential obstruction of the output oxygen.
Rapid audible beeping alert (this alert is deactivated when accessories are connected). May be
associated with flow setting of 0.5 L/min or less.
Perfecto
AW - 300 W at 5 L/min, 280 W at 3 L/min
2
Perfecto
VAW - 320 W at 5 L/min
2
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)
Technical data
31