Herunterladen Inhalt Inhalt Diese Seite drucken

Environmental Conditions; Owner's Guide - Coltene BioSonics UC300 Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
4
Note:
Beaker holder may be used in con-
junction with large basket (Fig. 9) or twin
basket. Remove basket section inserts seat-
ed directly under beakers to provide space
for beakers to seat fully. Use caution when
removing basket if beakers are left in place.
B. Twin FingerGuard Basket
The UC311 Twin FingerGuard Basket pro-
vides you with the versatility of cleaning
two basket loads of instruments or basket
and beaker cleaning for prostheses (or
other small items). Each Twin FingerGuard
Basket comes with two (2) basket section
Inserts.
Use Directions:
Place Twin FingerGuard Baskets side by
side, or use one basket with beaker hold-
er (Fig. 9,10). Twin FingerGuard Baskets
may be used with or without basket
section Inserts. Place instruments across
ridges in bottom of basket. Place basket
in tank and clean for desired time.
C. Basket Section Inserts
The UC335 basket section inserts keep
instruments separated by user and proce-
dure, eliminating sorting. This increases
efficiency and reduces the potential for
accidental sticks and cuts since instru-
ments are pre-sorted and handled less.
Use Directions:
Place locking tabs of adjustable inserts
into basket holes (as shown in Fig. 11).
Inserts can be placed at any desired width
along the basket. Place as many basket
section inserts as required for your clean-
ing needs. To remove or adjust position,
simply squeeze center tabs to unlock sec-
tion insert from basket holes.
D. Beaker Holder:
Place the UC330 beaker holder on bezel.
Place beaker positioning rings around the
beaker so that at least 2/3 of the beaker
is suspended in the main bath (Fig. 9).
Clean for desired time.
E. Cassette Racks:
Your BioSonic
UC300 has two instru-
®
ment cassette racks to choose from:
• UC325 Hu-Friedy IMS
/
1
Thompson
Cassette Rack
2
This rack will hold three (3) large or six (6)
small cassettes.
Use Directions:
Up to three (3) large cassettes may be
placed horizontally in each of the three
(3) tiers of the rack (Fig. 12). Up to six (6)
small cassettes can be placed either hor-
izontally or vertically in the basket (Fig.
13). Place filled basket in tank contain-
ing BioSonic
General Purpose Cleaner,
®
BioSonic
Enzymatic Cleaner or BioSonic
®
Germicidal Cleaner. Once cassettes are
cleaned, remove rack from tank and tilt
back over tank to drain. Rack has stops
on the back to prevent the cassettes from
slipping out (Fig. 14). Rinse rack and
cassettes on all sides thoroughly under
running tap water. Place the basket on
draining tabs of unit bezel to dry. (See
Instrument Draining section for more
details.)
Use Directions:
Place up to five (5) Steri-Instrument cas-
settes vertically into rack and place rack
into tank for cleaning (Fig. 15). For Zirc
or Clive Craig
instrument boxes, place
4
horizontally in each of the two (2) lower
tiers of the rack. Up to ten (10) instru-
ments boxes may be placed in the rack
(Fig. 16). Place rack in tank for cleaning.
After cleaning, rinse cassettes in rack
thoroughly under running tap water and
place on draining tabs of bezel to dry.
(See Instrument Draining section for more
details.)
Registered trademark of Hu-Friedy Inc.
1
Registered trademark of Thompson Inc.
2
Registered trademark of Zirc Inc.
3
Registered trademark of Clive Craig Inc.
4
V. Environmental Conditions
• Intended for indoor use
• Maximum altitude 2000m
• Temperature range:
40°C (104°F)
5°C (41°F)
• Humidity Range
80%@31°C (88°F)
50%@40°C (104°F)
• Mains supply voltage fluctuations not to
exceed ± 10% of the nominal voltage
• Transient over voltages category II
• Pollution degree 2
• Class I equipment
VI.
Cautions and Warning
NOTE: Failure to comply with these Caution
and Warning statements may result in injury
to personnel or equipment damage.
®
Operator's Guide
Do not operate the unit
without sufficient solution
in the tank.Do not use
water alone as the carrier
bath. Water is not a satisfac-
tory coupling agent unless a
wetting agent is used. For
best results use BioSonic
General Purpose, BioSonic
Germicidal
BioSonic
Ultrasonic Cleaning solu-
tions.
Do not place objects direct-
3
ly on bottom of the tank
because they may block cav-
itation and cause buildup
CAUTION!
of heat in the transducer
causing premature failure.
Use only BioSonic
accessory baskets, cassette
racks or beakers to hold
items.
Do not restrict airflow.
Provide
tilation by maintaining at
least 2 inches clearance at
the back.
Do not subject unit to shock
or impact.
Do not immerse the unit in
water.
This product is intended for
use only by trained health-
care professionals.
Use this product in accor-
dance with the instructions
in the owner's guide.
Unspecified or improper use
of this product may impair
safety protection.
Do not place fingers or
hands in the ultrasonic tank
while in operation.
Before cleaning items of
CAUTION!
considerable value ultra-
sonically, consult the item's
manufacturer for cleaning
precautions.
The power off selection
from the keypad does not
turn off main power to the
unit.
Never drain tank by manual-
ly discarding solution as this
may cause the unit to stop
working properly.
®
®
Cleaner
or
Enzymatic
®
®
UC300
adequate
ven-

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis