Conductivity system messages
System message
/
Indication range
exceeded
Zellenkonstante
außerhalb des
zulässigen Bereichs
General system messages
System message
Meter does not
react on
keypressing
LoBat
*(
is flashing
56
Cause
Measured value is outside of
measuring range
(Conductivity over 500 mS/cm)
Measuring cell contaminated
Unsuitable calibration
solutions
Cause
Undefined operation state
of the processor, e. g.
after inadmissible EMC
admission (outside the
specified ranges)
Rechargeable batteries
nearly discharged.
Timeout of serial interface.
Set calibration interval has
run out.
Troubleshooting
Action
Clean measuring cell, if
necessary exchange it
Check calibration solutions
Action
Reset of processor:
Press RCL key and
switch meter on.
Charge rechargeable batteries.
Check connected
instrument.
Recalibrate probe.