Measuring principle
Temperature compensation
Measuring conditions
Meas. range
Temperature range
Max. admissible
overpressure
Immersion depth
Operating position
Incident flow
Connections
Characteristics when delivered
Zero signal
Response time at 20°C
Self consumption at 20°C
Drift
Operation time
Material
Dimensions
90

Technical data

Membrane covered galvanic probe
IMT
0 to 50 mg/l O
2
0 °C to 50 °C
6 bar
min. 6 cm
max. 20 m depth
any
> 3 cm/s at 10 % meas. accuracy
10 cm/s at 5 % meas. accuracy
18 cm/s at 1 % meas. accuracy
Cable, plug: Protection degree IP67, 8pin plug
Probe:
Protection degree IP68 (20 m)
< 0.1 % of saturation value
t
(90 % of indication of end value after) < 10 sec
90
t
(95 % of indication of end value after) < 16 sec
95
t
(99 % of indication of end value after) < 60 sec
99
-1
0.008 µg h
(mg/l)
approx. 3 %/month in operation condition
at least 6 months per electrolyte filling
Membrane head
Membrane
Probe head
Thermistor housing
Shaft
shaft length
shaft diameter
membrane thickness 13 µm
cable length
-1
POM
FEP
POM
VA-steel 1.4571
POM
145 mm
15.25 mm
1.5 m to 20 m