General system messages
System message
Meter does not
react on
keypressing
LoBat
*(
is flashing
Troubleshooting
Cause
Undefined operation state
of the processor, e. g.
after inadmissible EMC
admission (outside the
specified ranges)
Batteries nearly discharged.
Timeout of serial interface.
Set calibration interval has
run out.
Oxi 330
Oxi 340
Action
Reset of processor:
Press RCL key and
switch meter on.
Charge rechargeable batteries/
Change non-rechargeable
batteries
Check connected
instrument.
Recalibrate probe.
45