In order to avoid instrument damage, it is ad-
vised to remove batteries when not using the
instrument over a certain period of time. Ho-
wever, should the instrument be contaminated
by leaking battery cells, you are kindly reque-
sted to return it to the factory for cleaning and
inspection.
Instruments must be stored in dry and enclosed
areas. In the case of an instrument being transpor-
ted in extreme temperatures, a recovery time of mi-
nimum 2 hours is required prior to instrument ope-
ration.

3.0 Safety Measures

The UNITEST TELARIS SCHLEIFE has been desi-
gned and checked in accordance with the safety re-
gulations for Electronic Test and Measurement In-
struments EN 61010 and IEC 1010, and left our
factory in a safe and perfect condition. To maintain
this condition the user must pay attention to the sa-
fety references contained in this instruction manual.
The respective accident prevention regulati-
ons established by the professional associati-
ons regarding electrical systems and equip-
ment must be strictly met for all tasks.
In order to avoid electrical shock, the valid sa-
fety and VDE regulations regarding excessive
contact voltages must receive the utmost at-
tention, when working with voltages excee-
ding 120V (60V) DC or 50V (25V)rms AC. The
values in brackets are valid for limited ranges
(eg for medicine and agriculture).
Safety Measures
If the operator's safety can no longer be gua-
ranteed, the instrument is to be put out of ser-
vice and protected against use. The safety is
no longer guaranteed, if the instrument:
• shows obvious damage
• does not carry out the desired measurement.
• has been stored for too long under unfavou-
rable conditions.
• has been subjected to mechanical stress du-
ring transport.
The instrument may only be used within the
operating ranges as specified in the technical
data section.
Avoid any heating up of the instrument by di-
rect sunlight to ensure perfect functioning and
long instrument life.
The Earth Loop Resistance Measurement In-
strument complies with the overvoltage class
CAT III / 300V and may only be used for mains
systems of maximum 230 V (fused at maxi-
mum 16 A).
Only use the mains cable supplied. The unit
was calibrated with this cable and only in this
way can the resistance values measured be
exactly indicated.