Herunterladen Inhalt Inhalt Diese Seite drucken

WIKA T16.H Betriebsanleitung Seite 28

Digitaler temperaturtransmitter
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

11. Specifications
11. Specifications
DANGER!
Danger to life due to loss of explosion protection
The non-observance of the instructions for use in hazardous areas can lead
EN
to the loss of the explosion protection.
Specifications
Permissible ambient temperature range
Climate class per IEC 654-1:1993
Maximum permissible humidity
Model T16.H
per IEC 60068-2-38:2009
Model T16.R
per IEC 60068-2-30:2005
Vibration resistance
per IEC 60068-2-6:2008
Shock resistance
per IEC 68-2-27:2009
Salt fog
per IEC 68-2-52:1996, IEC 60068-2-52:1996
Condensation
Free fall
in line with IEC 60721-3-2:1997,
DIN EN 60721-3-2:1998
Electromagnetic compatibility (EMC)
per DIN EN 55011:2010, DIN EN 61326-2-3:2013,
NAMUR NE21:2012, GL 2012 VI Part 7
{ } Items in curved brackets are options for an additional price, not for ATEX versions of the head mounting version and not for
T16.R rail mounting version
28
Adhere to the following limit values and instructions.
Model T16
{-50} -40 ... +85 {+105} °C
{-58} -40 ... +185 {+221} °F
Cx (-40 ... +85 °C / -40 ... +185 °F, 5 ... 95 % r. h.)
Test max. temperature variation 65 °C (177 °F) /
-10 °C (-18 °F), 93 % ±3 % r. h.
Test max. temperature 55 °C (131 °F), 95 % r. h.
Test Fc: 10 ... 2,000 Hz; 10 g,
amplitude 0.75 mm (0.03 in)
Acceleration / Shock width
Model T16.H: 100 g / 6 ms
Model T16.R: 30 g / 11 ms
Severity level 1
Model T16.H: Acceptable
Model T16.R: Acceptable in vertical mounting
position
Drop height 1.5 m (4.9 ft)
EN 61326 emission (group 1, class B) and
interference immunity (industrial application)
[HF field, HF cable, ESD, Burst, Surge]
WIKA operating instructions model T16