6. Commissioning, operation
The electrical connection is to be made according to the sensor connections/pin
assignments shown below:
Thermocouple with terminal block
GB
Single thermocouple
The colour coding at the positive connection to the devices always
decides the correlation of polarity and connection terminal.
Colour code of cable/terminal block
Sensor type
K
J
E
T

6.2 Cable connections

The cable gland must be optimally sealed, to ensure that the required protection class is
reached.
Requirements for meeting ingress protection
Only use cable glands within their indicated clamping range (cable diameter suitable for
the cable gland)
Do not use the lower clamping area with very soft cable types
Only use circular cross-section cables (if necessary, slightly oval in cross-section)
Do not twist the cable
Repeated opening/closing is possible; however only if necessary, as it might have a
detrimental effect on the protection class
For cable with a pronounced cold-flow behaviour the screw connection must be fully
tightened
CAUTION!
The protection class is not valid with armoured cables (stainless steel
sheathed)
Seal should be checked for signs of brittleness and, if necessary, replaced
10
Dual thermocouple
Standard
Positive
DIN EN 60584
green
DIN EN 60584
black
DIN EN 60584
violet
DIN EN 60584
brown
Triple thermocouple
Negative
white
white
white
white
WIKA operating instructions model TC90