1. On 77301 connect Motor Cable plug to mating receptacle on the Controller.
2. Mount pump head and load tubing. (See pump head manual.)
3. Connect power cord to Controller and grounded power line outlet.
4. Turn pump on and select tubing SIZE.
NOTE:
If CAL LED is lit, that tubing size has been previously field calibrated. If LED is not lit, the drive is operating
with the built-in factory calibration. To clear a field calibration, press and hold the CAL switch until the CAL
light goes out. This will take about 3 seconds. To recalibrate for better accuracy, see Calibration section.
5. MODE selection (INTernal, mA, or V).
6. Select pump DIRection (CW or CCW).
7. PRIME and CALibrate the pump (if required).
8. Press FLOW key and watch display to set the flow rate with UP/DOWN (▲, ▼) arrow keys.
9. Press START/STOP key to begin pumping.
NOTE:
When in INTernal mode, pump will not restart automatically after a brownout or powerout condition unless
operator changes default condition. In mA or V modes, pump will start with a non-zero mA or V signal.
Automatic Start Enable/Disable (INTernal mode only)
Press and hold START/STOP on power-up. After five (5) seconds, display will change to all dashes. Then, while
holding START/STOP, press PRIME five (5) times. Display will flash "ON" or "OFF". Use UP/DOWN (▲, ▼) arrow
keys to set automatic start option. Press any other key to exit. When "ON" is selected, pump will start automatically
at power-up if it was "ON" when powered down.

START/STOP Input for Models 77301-20 and 77301-30

This feature only applies to firmware version 3.01 or higher. To determine the firmware version of the drive, hold
the TUBING SELECT key when the drive is powered up. The firmware version number will appear on the display.
The START/STOP input can be configured to be optional (Factory Default), or required for the drive to run. It can
also be configured differently for Internal or External (mA or V) modes. For example, the START/STOP input
could be optional in Internal mode and required in External mode. To change the START/STOP input
configuration, first select the mode for the desired configuration (Internal, mA or V). Next press and hold the
MODE key until the display changes to "OFFSET" or "OFSET". Release the MODE key and press it again to skip
over the Offset setup. Now the display should be alternating between "STOP" and "OFF" or "ON". Next use the
"INC/DEC" (▲, ▼) arrow keys to select "ON" for required START/STOP input or "OFF" for an optional
START/STOP input. Press the MODE key again to save the setup. If you will be using both Internal and External
modes, be sure to perform the START/STOP setup in both modes.
With the "STOP OFF" option selected (Factory Default), use of the START/STOP input is optional. When the
START/STOP input is open, the drive can still be started using the START/STOP key, PRIME key, or PRIME
input. In external modes the drive will also run if there is sufficient current or voltage at the input. Closing the
START/STOP input will cause the drive to run until the START/STOP input opens or the START/STOP key is
pressed. In Volume Dispense, Copy Dispense, and Seconds mode, only a momentary START/STOP closure is
needed to star t the drive. If the drive is already running in one of the dispense modes, a momentary
START/STOP closure will stop the drive. In Set Copy mode, the START/STOP input functions the same as in
normal mode; closing it will cause the drive to run until it opens.

SETUP AND DRIVE OPERATION

2