10 Repairs

The sensor may only be repaired at the factory. Please return in its original
packaging. Prior to return, please carry out instructions as described in 11

"Decommissioning".

11 Decommissioning
Decommissioning sensor: observe all safety guidelines as given in section 5
"Assembly"!

12 Disposal

Sensor
IMPORTANT
Please observe the applicable national regulations.
You may return the decontaminated used equipment to ProMinent
Dosiertechnik GmbH, Heidelberg, against prepaid postage.
®
ProMinent
Repairs / Decommissioning / Disposal
Disconnect sensor from power supply (see 6 "Installation")
Depressurise the in-line probe housing
Loosen the locking screw
Slowly remove the sensor from the in-line probe housing
Unscrew the membrane cap over a wash basin or similar and empty
Rinse the membrane and electrodes with clean water and allow to dry (do
not allow dust to settle on electrodes or membrane)
Screw on a fresh membrane cap loosely to protect the electrodes
Fit the membrane cap cover to protect the membrane cap