 Once the step function is defi ned, set the switch
to Adj. Step and select Step Function n°3 using
the
or
key.
 Start the measurement by pressing START/
STOP
 During the measurement, the following screens can be accessed by pressing the DISPLAY key.
234.5 MΩ
999 V
224.6 pA
Remainning Time 00:14:30
1
1
1
10 100
10
100
10
100
 At the end of the measurement, the following results are indicated:
the difference   R in insulation resistance between the resistance at the highest test voltage
and the resistance at the lowest test voltage,
the difference   V between the highest and lowest test voltage,
the slope of the curve in ppm/V,
the capacitance.
 Pressing the GRAPH key displays the resistance versus time curve.
Using the
and
keys, it is possible to move along the curves and view the exact values of each
sample.
234.5 MΩ
999 V
Remainning Time 00:14:15
ΔR
ΔV
ΔR/(R*ΔV) (ppm/V)
10
Capacitance
95
STEP FUNCTION 3
Min: 1000 V
Test Run Time 00:15:00
Input voltage
Frequency
Input current
Date: 24.06.2008
224.6 pA
--- MΩ
300 V
---
---
Max: 5000 V
0.1 V DC
0.0 Hz
24.6 nA
Time: 15:31