Herunterladen Inhalt Inhalt Diese Seite drucken

Chauvin Arnoux C.A 6549 Bedienungsanleitung Seite 80

Megohmmeter
Vorschau ausblenden Andere Handbücher für C.A 6549:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

After the measurement
234.5 MΩ
2307 V
24.6 pA
Elapsed Time 00:10:00
DAR (30s/60s)
PI (1.0m/10m)
Capacitance
320 nF
Available information:
First screen
Insulation resistance
Measured voltage
Measured current
Test duration
DAR, PI, capacitance
 TIMED RUN + DD mode
Before the measurement
DD
ADJUSTABLE VOLTAGE 2
2300 V
Test Run Time 00:10:00
Input voltage
0.1 V AC
Frequency
0.0 Hz
Input current
24.6 pA
Date: 24.06.2008
Time: 15:32
Available information:
First screen
Selected test voltage
Programmed test duration
Input voltage
Frequency
Input current (DC only)
Date, time
ADJUSTABLE VOLTAGE 2
2300 V
Test Run Time 00:10:00
2.64
Input voltage
1.05
Frequency
Input current
Date: 24.06.2008
Press on DISPLAY
Selected test voltage
Programmed test duration
Input voltage
Frequency
Input current (DC only)
Date, time
DD
0.1 V
0.0 Hz
AC
V
0
250
500
100
Press on DISPLAY
Input voltage
Frequency
Input current (DC only)
Voltage bargraph
80
0.1 V
0.0 Hz
0.1 V AC
0.0 Hz
0
250
100
24.6 pA
Time: 15:32
2
nd
press on DISPLAY
Input voltage
Frequency
Input current (DC only)
Voltage bargraph
24.6 pA
750
1000
24.6 pA
AC
V
500
750
1000

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis