Temperature Unit: selection of temperature unit.
Default Probe Temperature: measurement temperature.
Rc Reference Temperature: reference temperature to which the measurement result must be
referred.
  T for R/2: estimated  T to obtain an insulation resistance / 2.
Calculate   T from Memory: used to calculate   T from 3 stored measurements made using the
same probe at different temperatures (see § 4.5.3).
Maximum Output Voltage: imposes maximum/locking of test voltage.
Set Default Parameter: default confi guration: reinitializes the instrument with the default values of
all parameters.
Clear Memory: can be used to partially or completely erase stored data (see § 4.5.2).
V Disturbance / V Output = dlSt factor (see § 3.2 - Important remark).
Buzzer: enabling / disabling of buzzer (keys, measurements, alarms).
Default value
°C
Default value
23 °C
Default value
40 °C
Default value
10 °C
Default value
5100 V
Default value
3%
Default value
On
88
Range
°C or °F
Range
-15°C ... +75°C
Range
-15°C ... +75°C
Range
-15°C ... +75°C
Range
40 ... 5100 V
Range
3, 10 or 20 %
Range
On or Off