Herunterladen Inhalt Inhalt Diese Seite drucken

Satelec Suprasson P5 Booster Benutzerhandbuch Seite 7

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 35
b) Rear of unit (Fig. 2)
• Water inlet tube (Fig. 2-1).
• Water filter (Fig. 2-2).
• Footswitch cord(Fig. 2-3).
• Mains power cable with functional earth
terminal (according to model) (Fig. 2-4).
• Tap for adjusting the water flow (Fig 2-5).
c) Technical Data
Power supply : 110 V or 220/230 V, 50/60 Hz
Class II equipment, BF Type
Intermittent operation 5 min / 10 min
Max power consumption : 40 VA
Vibration frequency : 27 to 33 kHz
Water pressure : 1 to 5 bar (14.5 to 73 psi)
Water filter : approx. 60 µm
Overall dimensions :
Depth : 150 mm
Width : 155 mm
Height : 60 mmWeight : 1.2 kg
Manufactured
in
compliance
regulations and standards in force (Directive
93/42/EEC CE medical devices).
1. 3 INSTALLATION
a) Recommendations
The use of the Suprasson P5 B
to competent and qualified practitioners (dental
surgeons, hygienists).
Note: before using for the first time, clean,
decontaminate and sterilize the accessories.
Before using the unit for the first time you are
recommended to plug in the accessories and
operate the spray for several minutes in order to
rinse the equipment's internal water circuits.
Check that the equipment is correctly plugged in
and insulated; the mains plug with earth pin must
be used.
The user must check the accessories and cords
regularly for insulation faults and replace them if
necessary.
Do not use the equipment if it appears to be
faulty.
If the equipment develops a fault you are
recommended to contact the supplier of your
Suprasson P5 B
specialist repairer.
b) Connections
The electrical connexion must conform to current
norms.
Rear panel of Suprasson P5 B
Connect:
• The flexible green water tube (Fig. 2-1) to the
mains water supply.
• The mains power cable should be plugged into
a earthed power supply (according to country).
with
the
c) Commissioning and use
Connecting the Suprasson handpiece.
Connecting the handpiece to the cable connector
is by a simple push fit. Do not twist.
Mounting and using the tips
To obtain adequate vibration of a tip it must be
is restricted
correctly tightened and not forced beyond its
OOSTER
stop. To avoid it becoming locked, the tip should
be removed after each use (Fig. 3).
See the TipBook for the power settings for
SATELEC tips.
d) WARNING
It is essential to turn off the water supply
before leaving the dental surgery.
7
rather than using a non-
OOSTER
(Fig. 2)
OOSTER

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis