Herunterladen Inhalt Inhalt Diese Seite drucken

Paragraph 3 - Timed Dosage (Frequency Signal Input "Trigger" Not Activated) - Oase Clear Fountain System CFS 50434 Gebrauchsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

- GB -
Pos: 79 /Gartentechnik/KDS Anlage TEKNA/Absatz 3 – Getaktete Dosierung (Eingang Frequenzsignal "TRIGGER" nicht aktiviert) @ 12\mod_1253631394014_121.doc @ 75049 @ 2
8.
Paragraph 3 – Timed Dosage (Frequency signal input "TRIGGER" not activated)
Programming
PROG
Configuration
Configuration
Pum functions
-
+
Pum functions
Tempo
Time
-
Quantity
100ml
Time
g.hh.mm
-
Delay
0.01.50
Time
g.hh.mm
-
Interval
0.01.50
Trigger Mode
-
Disable
Pause Input Mode
-
Restart Timer
Display during Operation
Operating mode
Timer
Alarms and Level
statuses = level alarm
Flow = flow alarm
Pos: 80 /Alle Produkte/Dummy_module/===== Seitenwechsel ===== @ 0\mod_1125648978015_0.doc @ 1631 @
30
-
+
-
+
-
+
-
+
Pause Input Mode
-
Restart Timer
Pause Input Mode
-
Freeze Time
Pause Input Mode
Pause Dosing
FLOW
sensor
status
Timer
7.0 ml
F
F F
Liv
Stop
P 0%
Pump status
Dry = pump starting up
Stop = pump stopping
Restart = pump restarting
Operation
The pump doses a programmed quantity in ml. It is possible to set a
pump delay time (Delay) when the pump is started and an interval
between two successive dosings (Interval), as illustrated in the
diagram:
Delay
The Delay and Interval times are in dd.hh.mm (days, hours,
minutes)
The Pause input can be programmed in three different modes:
FreezeTime: when the pause is activated, the system cuts out
the current time count and restarts it when the pause is
deactivated.
Pause Dosing: with the pause activated, the system continues
to count time the and stops the dosing.
Restart Timer: when the pause is activated, the system stops
the dosing and when the pause is deactivated the count starts
again from the beginning.
The dosage frequency can be modified while the pump is operating,
by pressing the
frequency, or the
Display at start-up (MODE key)
Modalità di funzionamento
Programmed value
Vengono visualizzati in
Quantity in ml
sequenza:quantità da
dosare, ritardo in g.hh.mm e
intervallo in g.hh.mm
Q.tà
Value of dosing in
20,0ml
progress:
Percentage of
Valore della percentuale di
strokes/min max
dosaggio, modificabile premendo
sui tasti + o -
Quantity
Interv.
keys at the same time to increase the
keys to decrease it.
Valore del dosaggio
Quantità
20,0ml
programmata in
P 100%
ml

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis