ZERO_MEAS?
Liefert den Offsetnulldruck bzw. den Referenzwert für Absolutdruckmodule zurück.
8. Specifications
All measurements apply at 23°C ± 5°C. unless specified otherwise. Outside of this range the
stability of the measurements is ± 0.005%of reading/°C.
Operating Temperature
Storage Temperature
Power
Low battery warning
Serial Communications
CE - EMC
Safety
Read: Isolated(Upper Display)
Read: non-Isolated(Lower Display)
Source
Maximum current output in voltage ranges is 3mA with an output impedance of <= 1W.
Read: Isolated(Upper Display)
Read: non-Isolated(Lower Display)
Source
Maximum load on mA source is 1000W. Voltage input range on simulate mode 5V - 30V.
Read
Source
Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude is adjustable
from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight negative offset of
approximately -0.1V is present to assure zero crossing.
Ohms low
Ohms high
Range
Ohms low
5.0W - 400.0W
Table 9: General Specifications
-10°C to 50°
-20°C to 70°C
4 X AA batteries; Alkaline or optional rechargeable
Yes
Yes, ASCII
EN50082-1: 1992 and EN55022: 1994 Class B
CSA C22.2 No. 1010.1: 1992
Table 10: DC Voltage Measurement/Source
Range
0.000V- 30.000V
0.000V - 20.000V
0.000V - 20.000V
Table 11: DC mA Measurement/Source
Range
0.000mA - 24.000mA
0.000mA - 24.000mA
0.000mA - 24.000mA
Table 12: Frequency Measurement/Source
Range
2.0CPM - 600.0CPM
1.0Hz - 1000.0Hz
1.00KHz - 10.00KHz
2.0CPM - 600.0CPM
1.0Hz - 1000.0Hz
1.00KHz - 10.00KHz
Table 13: Resistance Measurement
Range
0.00W - 400.0W
401.0W - 4000.0W
Table 14: Resistance Source
Excitation Current
0.1mA - 0.5mA
Accuracy (% of reading ± floor)
0.015% ± 2mV
0.015% ± 2mV
0.015% ± 2mV
Accuracy(% of reading ± floor)
0.015% ± 2mA
0.015% ± 2mA
0.015% ± 2mA
Accuracy(% of reading ± floor)
0.05% ± 0.1CPM
0.05% ± 0.1Hz
0.05% ± 0.01KHz
0.05%
0.05%
0.125%
Accuracy(% of reading ± floor)
0.015% ± 0.03W
0.015% ± 0.3W
Accuracy(% of reading ± floor)
0.015% ± 0.1W